Battery Insertion

1. Remove the battery door by sliding it out of place.

2. Insert the 2 included AAA batteries. Note: They insert the same way.

3. Slide the battery cover back on. Note: It can be finicky and might get stuck, but be gentle and it will insert and click into place. 



Taking apart the spirometer

1. Gently turn the the orange turbine counterclockwise and then slide the turbine and mouth piece out.

2. Gently pull the mouthpiece out of the turbine.


To wash the turbine and mouthpiece

 

Turbine

Mouthpiece

Handheld piece

To avoid irreparable damage to the turbine, do not use any alcoholic or oily cleaning solutions, and do not immerse in hot water or other solutions. Do not run under a direct jet of water or other solution


Submerge in warm soapy water for 2-3 minutes. Rinse in clean water and shake gently to remove excess water. let the turbine dry completely before next use

Submerge in warm soapy water for 2-3 minutes. Rinse in clean water and shake gently to remove excess water. let dry completely before next use



The mouthpiece should be cleaned after each use


Do not ever submerge this piece in water or any other liquidWipe down with a damp cloth and allow surfaces to air dry


To re-assemble the spirometer 

1. Replace the mouthpiece back into the turbine.

2. Line up the turbine with the spirometer body, orange side to orange side. 

3. Insert and give a quarter turn clockwise to ensure the turbine locks into place.